Philips North America recalls (1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.)
- Products
-
- (1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)
- Company
- Philips North America
- Where it was sold
-
Domestic distribution to the following states and territories: CA DE FL GA IL KY MA ME MI MO NH NJ NM NY OH PA SC TN TX VA VT WI WY, and PR. International distribution worldwide.
States: California, Delaware, Florida, Georgia, Illinois, Kentucky, Massachusetts, Maine, Michigan, Missouri, New Hampshire, New Jersey, New Mexico, New York, Ohio, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, Virginia, Vermont, Wisconsin, Wyoming
- Barcodes
0088483809004000884838090057- Units affected
- 56 US; 169 ROW
- About this recall
- (1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)