Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips North America recalls (1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The sealed magnet will experience uncontrolled shutdown known as Loss-of-Field (LoF) if the magnet loses cryo-cooling (e.g. caused by a power outage on site.)
Products
  • (1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)
Where it was sold

Domestic distribution to the following states and territories: CA DE FL GA IL KY MA ME MI MO NH NJ NM NY OH PA SC TN TX VA VT WI WY, and PR. International distribution worldwide.

States: California, Delaware, Florida, Georgia, Illinois, Kentucky, Massachusetts, Maine, Michigan, Missouri, New Hampshire, New Jersey, New Mexico, New York, Ohio, Pennsylvania, Puerto Rico, South Carolina, Tennessee, Texas, Virginia, Vermont, Wisconsin, Wyoming

Barcodes
0088483809004000884838090057
Units affected
56 US; 169 ROW
About this recall
(1) Ingenia Ambition X (Product Number 781356); (2) Ingenia Ambition S (Product Number 781359)