Trokamed GmbH recalls Sheath
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure.
- Products
-
- Sheath
- Company
- Trokamed GmbH
- Where it was sold
-
US (Domestic) distribution in state of PA and OUS (foreign) to country of Germany. .
States: Pennsylvania
- Barcodes
04251303810865- Units affected
- 53 units
- About this recall
- Sheath, 18 Fr., 130 mm (REF: WA2PS18S), gastroenterology-urology endoscopic access overtube.