Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Sarnova HC, Llc recalls Curaplex Field Cric Kit with ET Tube

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.
Products
  • Curaplex Field Cric Kit with ET Tube
Where it was sold

US Nationwide

Nationwide

Units affected
1063 kits
About this recall
Curaplex Field Cric Kit with ET Tube, Item Number 36611. The kit provides healthcare professionals with an assortment of supplies to perform a cricothyrotomy.