Medtronic Inc. recalls Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the test screen did not prevent retesting after a failure. Some batteries were therefore retested and passed.
Products
  • Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator
Where it was sold

US Nationwide distribution in the state of Florida.

States: Florida

Barcodes
00763000178338
Units affected
1 device
About this recall
Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator, product number DDPA2D1