Medtronic Inc. recalls Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the test screen did not prevent retesting after a failure. Some batteries were therefore retested and passed.
- Products
-
- Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator
- Company
- Medtronic Inc.
- Where it was sold
-
US Nationwide distribution in the state of Florida.
States: Florida
- Barcodes
00763000178338- Units affected
- 1 device
- About this recall
- Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator, product number DDPA2D1