Beckman Coulter, Inc. recalls Beckman Coulter DxU 850m/840m Microscopy Analyzers
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.
- Products
-
- Beckman Coulter DxU 850m/840m Microscopy Analyzers
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
US and Australia, Brazil, Canada, China, France, Germany, Ghana, India, Israel, Italy, Lebanon, Lithuania, Mexico, Panama, Qatar, Romania, Saudi Arabia, Spain, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland,
- Barcodes
1509959074864715099590748654- Units affected
- 254 units
- About this recall
- Beckman Coulter DxU 850m/840m Microscopy Analyzers, Part Numbers: C49513, C76947