Philips North America recalls Intera 1.5T

Class IIIofficial

FDA Class III

FDA classified this recall as Class III: the product is unlikely to cause health problems but breaks the agency's rules (a labeling mistake, for example).

Hazard
An incorrect electrical shock warning label may have been placed on the Liquid Cooling Cabinet (LCC) of affected MR systems. The caution symbol on the incorrect label has a white background instead of yellow, and the label is missing the high voltage warning symbol. This presents a potential risk of electrical shock for service personnel.
Products
  • Intera 1.5T
Where it was sold

Worldwide - US Nationwide distribution in the states of AR, CO, IL, MD, MS, NE, TX and the countries of Brazil, Canada, China, Germany, Greece, India, Mexico, and Saudi Arabia.

States: Arkansas, Colorado, Illinois, Maryland, Mississippi, Nebraska, Texas

Units affected
7 US; 16 ROW total
About this recall
Intera 1.5T, Model No. 781195