Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

XTANT Medical Holdings, Inc recalls Cortera Spinal Fixation System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.
Products
  • Cortera Spinal Fixation System
Where it was sold

US distribution in MI.

States: Michigan

Barcodes
00810076564005
Units affected
6
About this recall
Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.