XTANT Medical Holdings, Inc recalls Cortera Spinal Fixation System
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.
- Products
-
- Cortera Spinal Fixation System
- Company
- XTANT Medical Holdings, Inc
- Where it was sold
-
US distribution in MI.
States: Michigan
- Barcodes
00810076564005- Units affected
- 6
- About this recall
- Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.