Bard Peripheral Vascular Inc recalls Safe-T-Centesis 6 Fr Catheter Drainage Tray
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.
- Products
-
- Safe-T-Centesis 6 Fr Catheter Drainage Tray
- Company
- Bard Peripheral Vascular Inc
- Where it was sold
-
US Nationwide distribution in the states of CA, MN, IN, MI, LA.
States: California, Indiana, Louisiana, Michigan, Minnesota
- Units affected
- 1240
- About this recall
- Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,