Beckman Coulter, Inc. recalls Access 2 Reaction Vessels
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.
- Products
-
- Access 2 Reaction Vessels
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
US Nationwide. Global Distribution.
Nationwide
- Barcodes
15099590180249- Units affected
- 38,225 untis
- About this recall
- Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibration samples on the Access 2 Immunoassay system, Catalog Number 81901