Ge Medical Systems Israel, Functional Imaging recalls GE Healthcare Discovery NM/CT 670 CZT
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- Products
-
- GE Healthcare Discovery NM/CT 670 CZT
- Where it was sold
-
Worldwide
Nationwide
- Barcodes
840682121446- Units affected
- 8 units
- About this recall
- GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission