Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ge Medical Systems Israel, Functional Imaging recalls GE Healthcare Infinia I Hawkeye 1

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Products
  • GE Healthcare Infinia I Hawkeye 1
Where it was sold

Worldwide

Nationwide

Units affected
93 units
About this recall
GE Healthcare Infinia I Hawkeye 1, Model/Catalog Numbers: 1) H3000WN; 2) H3000YM; 3) H3000YS; System, Tomography, Computed, Emission