Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Quidel Ortho recalls Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Hemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively biased digoxin results.
Products
  • Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number
Where it was sold

Domestic: Nationwide Distribution International: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom,

Nationwide

Barcodes
10758750004782
Units affected
4,417 units (3,358 US, 1,125 OUS)
About this recall
Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VITROS DGXN Slide is a multilayered, analytical element coated on a polyester support. Digoxin is a widely prescribed cardiac glycoside indicated in the treatment of congestive heart failure and supraventricular arrhythmias. Digoxin measurements are used to monitor patient compliance and therapy, and to diagnose potential overdose. VITROS Chemistry Products DGXN Slides quantitatively measure digoxin (DGXN) concentration in serum using VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. One sales unit contains 5 cartridges. Each cartridge holds 18 slides (equivalent to one test), for 90 slides per sales unit. Component: No, not a component