Philips North America recalls BrightView X

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.
Products
  • BrightView X
Where it was sold

Domestic distribution nationwide. International distribution worldwide.

Nationwide

Units affected
56 US; 46 OUS
About this recall
BrightView X, gamma camera for SPECT; Model Nos.: 6-digit format 882478; 12-digit format 453560824741 453560829261;