Philips North America recalls BrightView XCT
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.
- Products
-
- BrightView XCT
- Company
- Philips North America
- Where it was sold
-
Domestic distribution nationwide. International distribution worldwide.
Nationwide
- Units affected
- 106 US; 241 OUS
- About this recall
- BrightView XCT, gamma camera for SPECT; Model Nos.: 6-digit format 882482; 12-digit format 453560462131 453560749161