Path recalls MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.
- Products
-
- MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)
- Company
- Path
- Where it was sold
-
Domestic: WI. International: Ireland.
States: Wisconsin
- Barcodes
04260223141355- About this recall
- MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)