Philips North America recalls Achieva 1.5T Conversion. Model (REF) Numbers 781346
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Products
-
- Achieva 1.5T Conversion. Model (REF) Numbers 781346
- Company
- Philips North America
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
0088483806846900884838009806- Units affected
- 7 US; 137 OUS
- About this recall
- Achieva 1.5T Conversion. Model (REF) Numbers 781346, 781283.