Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips North America recalls Achieva 1.5T Conversion. Model (REF) Numbers 781346

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
Products
  • Achieva 1.5T Conversion. Model (REF) Numbers 781346
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
0088483806846900884838009806
Units affected
7 US; 137 OUS
About this recall
Achieva 1.5T Conversion. Model (REF) Numbers 781346, 781283.