Philips North America recalls Achieva 1.5T Initial system. Model (REF) Numbers 781178.

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
Products
  • Achieva 1.5T Initial system. Model (REF) Numbers 781178.
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Units affected
23 US; 51 OUS
About this recall
Achieva 1.5T Initial system. Model (REF) Numbers 781178.