Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips North America recalls Achieva 3.0T. Model. (REF) Numbers 781277

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
Products
  • Achieva 3.0T. Model. (REF) Numbers 781277
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
00884838004139
Units affected
75 US; 442 OUS
About this recall