Philips North America recalls Ingenia 3.0T CX. Model (REF) Numbers 781271
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Products
-
- Ingenia 3.0T CX. Model (REF) Numbers 781271
- Company
- Philips North America
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
0088483806845200884838098312- Units affected
- 1 US; 347 OUS
- About this recall
- Ingenia 3.0T CX. Model (REF) Numbers 781271, 782105.