Philips North America recalls Intera 1.5T Achieva Nova. Model (REF) Numbers 781172.

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
Products
  • Intera 1.5T Achieva Nova. Model (REF) Numbers 781172.
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Units affected
12 US; 45 OUS
About this recall
Intera 1.5T Achieva Nova. Model (REF) Numbers 781172.