Philips North America recalls SmartPath to dStream for 1.5T. Model (REF) Numbers 781260
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
- Products
-
- SmartPath to dStream for 1.5T. Model (REF) Numbers 781260
- Company
- Philips North America
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
0088483809507600884838098886- Units affected
- 37 US; 735 OUS
- About this recall
- SmartPath to dStream for 1.5T. Model (REF) Numbers 781260, 782112.