Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Percussionaire Corporation recalls VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.
Products
  • VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK
Where it was sold

Worldwide - U.S. Nationwide distribution in the states of AZ, FL, IL, MS, MO, NE, NJ, NY, NC, RI, TX, UT, and WV. The countries of Belgium, Canada, France, Hong Kong, Netherlands, Russia, Switzerland and Turkey.

States: Arizona, Florida, Illinois, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, New York, Rhode Island, Texas, Utah, West Virginia

Barcodes
00849436000259
Units affected
4,727 packs/cases
About this recall
VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR W/SWIVEL T SINGLE PATIENT PHASITRON -PRT 992, VDR4 HUMIDIFICATION ADAPTER KIT, CASE OF 10 -PRT 993, VDR4 Humidification Kit with Cross Tee, Case of 10 The Phasitron breathing circuit kit is intended to be used for continuous ventilation intended for controlled ventilation of patients