Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

CareFusion 303, Inc. recalls BD Pyxis MedStation ES

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.
Products
  • BD Pyxis MedStation ES
Where it was sold

Worldwide - US Nationwide distribution including in the states of CO, IN, OH, TX, KS, FL, GA, NC, CA, HI, IL, NY, MI, VA, AL, LA, NE, KY, OK, AR, TN, WY, NJ, WI, SD, AZ, NV, MO, MA, ID, WV, DE, PA, WA, MT, CT, IA, MD, PR, MN, NM, OR, VT, MS, ME, AK, UT, DC, GU, SC, NH, RI, ND, AS and the countries of CH, CA, KW, QA, NL, GB, UY, AU, NZ, BH, TH, SG, CN, FR, LB, BS, HK, TR, PL, AR, BR, CL, CO, MX, ES, BM, SA, IL.

Nationwide

Units affected
53669
About this recall
BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower, REF: 352, BD Pyxis Pro 7-Drawer Auxiliary, REF: 1149-00, BD Pyxis Pro 7- Drawer Auxiliary with Secure Tower, REF: 1152-00, BD Pyxis Pro MedStation Main, REF: 1155-00, BD Pyxis CII Safe ES, REF: 1116-00