Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips North America recalls Ingenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
Products
  • Ingenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732)
Where it was sold

Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.

States: Florida, New York

Barcodes
00884838009820
Units affected
1
About this recall
Ingenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system