Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Baxter Healthcare Corporation recalls Baxter Exactamix Pro 2400

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
An error was identified in software versions 2.0.8 and 2.1.8 while using the "Use Some Overfill" feature which may lead to over-delivery of an ingredient. This issue can result in redundant ingredient delivery of the requested overfill volume.
Products
  • Baxter Exactamix Pro 2400
Where it was sold

US: AL, AZ, CA, FL, GA, IA, KY, MD, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OR, SD, TX, WA, WI,

States: Alabama, Arizona, California, Florida, Georgia, Iowa, Kentucky, Maryland, Michigan, Minnesota, Missouri, North Carolina, Nebraska, New Hampshire, New Jersey, Nevada, New York, Oregon, South Dakota, Texas, Washington, Wisconsin

Barcodes
05413765588136
Units affected
67 devices
About this recall
Baxter Exactamix Pro 2400, REF EXM24DY