Medline Industries, LP recalls MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
- Products
-
- MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed
- Company
- Medline Industries, LP
- Where it was sold
-
US, CA, PA, GUAM, VI
States: California, Guam, Pennsylvania, Virgin Islands
- Barcodes
40080196321002- Units affected
- 138412 units
- About this recall
- MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed, REF MDR107003E; 2) Full Electric Basic Homecare Bed, 4-pack,REF MDR107003E-4.