Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Products
  • MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed
Where it was sold

US, CA, PA, GUAM, VI

States: California, Guam, Pennsylvania, Virgin Islands

Barcodes
10080196546992
Units affected
22565 units
About this recall
MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed, REF MDR107003ELO