Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS)

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.
Products
  • Ivenix Infusion System (IIS)
Where it was sold

US distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI.

States: California, Colorado, Georgia, Idaho, Michigan, Mississippi, Nebraska, New Jersey, Nevada, Oklahoma, Texas, Utah, Virginia, Washington, Wisconsin

Barcodes
00811505030122
Units affected
23 systems
About this recall
Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.