Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

ROi CPS LLC recalls regard Item Number: 830106006

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..
Products
  • regard Item Number: 830106006
Where it was sold

Distribution was made to MO. There was no government/military/foreign distribution.

States: Missouri

Barcodes
10194717119074
Units affected
414 kits
About this recall
regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.