ROi CPS LLC recalls regard Item Number: 830106006
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..
- Products
-
- regard Item Number: 830106006
- Company
- ROi CPS LLC
- Where it was sold
-
Distribution was made to MO. There was no government/military/foreign distribution.
States: Missouri
- Barcodes
10194717119074- Units affected
- 414 kits
- About this recall
- regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.