Mirion Technologies (Capintec), Inc. recalls Captus 4000e Thyroid Uptake Systems- Intended to be used by trained Nuclear Medicine Technologists or Nuclear
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Spring arm failure with Captus 4000e Thyroid Uptake Systems, can cause the collimator to fall downward to its lowest point of travel, causing potential injury with a patient or operator
- Products
-
- Captus 4000e Thyroid Uptake Systems- Intended to be used by trained Nuclear Medicine Technologists or Nuclear
- Where it was sold
-
Nationwide Foreign: AUSTRIA BANGLADESH BRASIL Brazil CANADA CHILE Costa Rica Ecuador France Hanoi Vietnam Hong Kong INDIA ISRAEL ITALY Malaysia PAKISTAN Philippines POLAND Saudi Arabia SOUTH KOREA Spain SWEDEN TAIWAN THAILAND TURKEY UNITED ARAB EMIATES UNITED KINGDOM VIETNAM
Nationwide
- Barcodes
0859942006096- Units affected
- 467 units
- About this recall
- Captus 4000e Thyroid Uptake Systems- Intended to be used by trained Nuclear Medicine Technologists or Nuclear Medicine Physicians to perform Thyroid Uptake Procedures. Models: 5430-30151, 5430-30152, and 5430-30154