Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

LivaNova USA, Inc. recalls SenTiva

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
Products
  • SenTiva
Where it was sold

US: AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT. OUS: AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR, HR, IL, IN, IT, JO, KW, LB, MK, NL, OM, PA, PE, PL, QA, RO, RS, SA, SE, SG, SI, TR, TT, UK, ZA.

States: Alaska, Arizona, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, North Dakota, Nebraska, New Jersey, New Mexico, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia

Barcodes
05425025750405
Units affected
48
About this recall
SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimulation (VNS).