Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Microbiologics Inc recalls Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.
Products
  • Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet)
Where it was sold

US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NE, NH, NJ, NY, OH, PA, RI, TN, TX, VA, WA, WI, WV, and WY. There was government distribution and no military distribution. Foreign distribution was made to France.

States: Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Massachusetts, Michigan, Minnesota, Missouri, Mississippi, Montana, Nebraska, New Hampshire, New Jersey, New York, Ohio, Pennsylvania, Rhode Island, Tennessee, Texas, Virginia, Washington, Wisconsin, West Virginia, Wyoming

Barcodes
70845357041448
Units affected
580 packages
About this recall
Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.