Zimmer Surgical Inc recalls 3:1 Dermacarrier
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to manufacturing issue, the product is labeled as 3:1 ratio, but the ridge pattern is a 1.5:1 ratio. The problem may lead to minor tissue damage or additional unplanned graft if deemed necessary.
- Products
-
- 3:1 Dermacarrier
- Company
- Zimmer Surgical Inc
- Where it was sold
-
Worldwide - US Nationwide distribution including in the states of CA, DE, FL, GA, IA, IL, IN, KS, LA, MI, MO, MT, NC, NV, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI, WV and the countries of AUTRALIA, CANADA, DNIPRO OBLAS, HONG KONG, INDIA, JAPAN, MEXICO, CDMX, NETHERLANDS, SEOUL KOREA, SINGAPORE, TAIWAN, THAILAND.
Nationwide
- Barcodes
00889024378780- Units affected
- 76 boxes (1,520 units) US; 521 boxes (10,420 units) OUS
- About this recall
- 3:1 Dermacarrier, Model Number 00219501300, skin graft carrier