Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Elekta, Inc. recalls MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
Products
  • MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Where it was sold

US Nationwide distribution in the state of TX.

States: Texas

Barcodes
07340201500026
Units affected
1
About this recall
MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)