Medline Industries, LP recalls Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
- Products
-
- Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS
- Company
- Medline Industries, LP
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
1019845926919610198459326318- Units affected
- 96 units
- About this recall
- Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359