Medline Industries, LP recalls Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
- Products
-
- Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE
- Company
- Medline Industries, LP
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
10198459156991- Units affected
- 12 units
- About this recall
- Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G