Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Beckman Coulter Inc. recalls Access PCT Reagent Pack

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.
Products
  • Access PCT Reagent Pack
Where it was sold

Worldwide distribution - US Nationwide and the countries of Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Kazakhstan, Korea, Republic of, Malaysia, Mexico, Peru, Philippines, Poland, Puerto Rico, Romania, Slovakia, South Africa, Spain, Switzerland, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Viet Nam.

Nationwide

Barcodes
15099590736002
Units affected
6,994 US; 10,067 OUS
About this recall
Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems