Baxter Healthcare Corporation recalls SIGMA Spectrum Infusion Pump
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.
- Products
-
- SIGMA Spectrum Infusion Pump
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
00085412498683- Units affected
- 585 units
- About this recall
- SIGMA Spectrum Infusion Pump, Model/Catalog Number: 35700BAX2