Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Microbiologics Inc recalls Microbiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.
Products
  • Microbiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet)
Where it was sold

Worldwide distribution - US Nationwide and the countries of Netherlands, Sweden, UK, France, Australia, New Zealand.

Nationwide

Barcodes
70845357043077
Units affected
27 units
About this recall
Microbiologics Multi-Strain Human Papillomavirus (HPV) Verification Panel (Inactivated Pellet), REF 8235; in-vitro diagnostic test