Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls Medline medical procedure convenience kits labeled as: 1) DEPAUL BASIC NEURO-LF

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.
Products
  • Medline medical procedure convenience kits labeled as: 1) DEPAUL BASIC NEURO-LF
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
101984592289951019845952640410198459443138101984594954271019845923133910198459553660
Units affected
236 units
About this recall
Medline medical procedure convenience kits labeled as: 1) DEPAUL BASIC NEURO-LF, Kit SKU DYNJ20028L; 2) NEURO MINOR PACK, Kit SKU DYNJ46546M; 3) NEURO SHUNT PACK, Kit SKU DYNJ54396K; 4) NEURO PACK, Kit SKU DYNJ59397Q; 5) NEURO BASIC, Kit SKU DYNJ904052L; 6) PK NEURO SHUNT BUMC, Kit SKU DYNJT6428.