Intelerad Medical Systems Incorporated recalls IntelePACS (Image Fusion Module) - InteleViewer
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.
- Products
-
- IntelePACS (Image Fusion Module) - InteleViewer
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia
States: Alabama, Arkansas, Arizona, California, Colorado, Florida, Illinois, Indiana, Maryland, Michigan, Minnesota, North Carolina, Nebraska, New Hampshire, New Jersey, Ohio, South Carolina, Tennessee, Virginia, Washington, Wisconsin, West Virginia, Wyoming
- Units affected
- 88
- About this recall
- IntelePACS (Image Fusion Module) - InteleViewer