Ortho-Clinical Diagnostics, Inc. recalls VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.
- Products
-
- VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Nationwide
- Barcodes
10758750012343- Units affected
- 15 US; 43 OUS
- About this recall
- VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)