Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation System
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
- Products
-
- CellMek SPS Sample Preparation System
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.
States: Alabama, Illinois, Massachusetts, Missouri, Oregon, Washington
- Barcodes
15099590750312- Units affected
- 50
- About this recall
- CellMek SPS Sample Preparation System, REF: C44603, with software