Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
Products
  • CellMek SPS Sample Preparation System
Where it was sold

Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.

States: Alabama, Illinois, Massachusetts, Missouri, Oregon, Washington

Barcodes
15099590750312
Units affected
50
About this recall
CellMek SPS Sample Preparation System, REF: C44603, with software