Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Pentax of America Inc recalls PENTAX Medical Video Colonoscope- Intended to provide optical visualization of (via a video monitor)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.
Products
  • PENTAX Medical Video Colonoscope- Intended to provide optical visualization of (via a video monitor)
Where it was sold

Nationwide

Nationwide

Barcodes
04961333252169
Units affected
152 units
About this recall
PENTAX Medical Video Colonoscope- Intended to provide optical visualization of (via a video monitor), and therapeutic access to the lower gastrointestinal tract Model: EC38-i20cL