Pentax of America Inc recalls Pentax Medical Video Processor- Intended to be used with PENTAX Medical endoscopes
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.
- Products
-
- Pentax Medical Video Processor- Intended to be used with PENTAX Medical endoscopes
- Company
- Pentax of America Inc
- Where it was sold
-
Nationwide
Nationwide
- Barcodes
04961333247974- Units affected
- 94 units
- About this recall
- Pentax Medical Video Processor- Intended to be used with PENTAX Medical endoscopes, monitors and other peripheral devices for endoscopic diagnosis, treatment, and video observation. Model: EPK-i8020c