Micro Therapeutics, Inc. recalls The Apollo Onyx Delivery Microcatheter (AOMC) is a DMSO-compatible
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- European version of microcatheter were distributed within US which contain a different "Indications for Use".
- Products
-
- The Apollo Onyx Delivery Microcatheter (AOMC) is a DMSO-compatible
- Company
- Micro Therapeutics, Inc.
- Where it was sold
-
US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, HI, IA, IL, IN, KS, MA, MD, MI, MO, NC, ND, NE, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI.
States: Alabama, Arizona, California, Connecticut, Florida, Georgia, Hawaii, Iowa, Illinois, Indiana, Kansas, Massachusetts, Maryland, Michigan, Missouri, North Carolina, North Dakota, Nebraska, Nevada, New York, Ohio, Oklahoma, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin
- Barcodes
00763000311438- Units affected
- 82 units
- About this recall
- The Apollo Onyx Delivery Microcatheter (AOMC) is a DMSO-compatible, single-lumen, end hole catheter designed for the infusion of the Onyx liquid embolic system (LES). The catheter has a semi-rigid proximal shaft and a highly flexible distal shaft to facilitate the advancement of the catheter in the anatomy. The proximal end of the catheter incorporates a standard Luer adapter to facilitate the attachment of accessories. The outer diameter surface of the catheter is hydrophilically coated to increase lubricity. The AOMC is designed to facilitate catheter retrieval in the event the catheter becomes entrapped within the vasculature. The distal section of the catheter incorporates a detachment zone that allows detachment of the distal tip when the force required to extract the catheter exceeds the force to detach the tip. The catheter has two radiopaque marker bands to visualize the position of the catheter and the detach zone area, one proximal to the detach zone and one at the distal end of the catheter.