Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

ARROW INTERNATIONAL Inc. recalls ARROW QuickFlash Radial Artery Catherization (wire

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire handle/chamber during use. The possible immediate health consequences of the component issue are arterial vasospasm and vessel injury arising from multiple arterial punctures with repeated attempts.
Products
  • ARROW QuickFlash Radial Artery Catherization (wire
Where it was sold

Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: AE, AR, AT, AU, BE, BH, BR, CA, CN, CO, CY, DE, EC, EE, ES, FR, GR, HK, HR, IT, KW, LB, LU, MX, MY, NL, NZ, PH, PL, QA, RU, SA, SE, SG, SI, TH, TW, and UK.

Nationwide

Barcodes
20801902190943308019021216162080190211166520801902190851208019021564064080190212144620801902159124208019021140484080190211800220801902139416208019021366062080190218067830801902193309308019022106313080190219331630801902202988
Units affected
1,160,470 devices
About this recall
ARROW QuickFlash Radial Artery Catherization (wire, guide, catheter), Part Numbers: a) REF AK-04220; b) REF ASK-04220-FMH1; c) REF ASK-04220-HHC; d) REF ASK-04220-KSP; e) REF ASK-04220-PSU; f) REF ASK-04220-UCL1; g) REF ASK-04500-AH; h) REF ASK-04500-HF-S; i) REF NA-04220-S1A; j) REF NA-04220-X1A; k) REF RA-04220; l) REF RA-04220-W