Baxter Healthcare Corporation recalls STERIS Boot Assembly
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Three lots of the Boot Assembly, Traction Device are being recalled due to due to a customer report of an incorrectly oriented locking pin which caused the boot to detach from the traction system. If an impacted device is used during surgery, the boot may detach from the traction system, potentially leading to patient harm and a delay in surgery.
- Products
-
- STERIS Boot Assembly
- Where it was sold
-
US Nationwide distribution in the state of Alabama.
States: Alabama
- Barcodes
00615521000851- Units affected
- 14 units
- About this recall
- STERIS Boot Assembly, Traction Device, OT 1000, Part Number REF 10009192