Lumenis, Ltd. recalls (1) Lumenis Pulse 120H 30 amp
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.
- Products
-
- (1) Lumenis Pulse 120H 30 amp
- Company
- Lumenis, Ltd.
- Where it was sold
-
Distribution was made to PA. There was no government/military distribution. Foreign distribution was made to Mexico and Spain.
States: Pennsylvania
- Barcodes
072901091455180729010914552507290109145464- Units affected
- 3 devices
- About this recall
- (1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF GA0006802, Laser System; and (3) Lumenis Pulse 120H, REF GA-0008700, Laser System.