All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Lumenis, Ltd. recalls (1) Lumenis Pulse 120H 30 amp

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.
Products
  • (1) Lumenis Pulse 120H 30 amp
Where it was sold

Distribution was made to PA. There was no government/military distribution. Foreign distribution was made to Mexico and Spain.

States: Pennsylvania

Barcodes
072901091455180729010914552507290109145464
Units affected
3 devices
About this recall