Maquet Cardiopulmonary Gmbh recalls Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.
- Products
-
- Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;
- Company
- Maquet Cardiopulmonary Gmbh
- Where it was sold
-
Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea, Republic of (South Korea), Latvia, Lithuania, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Uzbekistan, and Vietnam
Nationwide
- Barcodes
04037691816432- Units affected
- 3050 units
- About this recall
- Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;